產(chǎn)品簡介
product
產(chǎn)品分類Primary and active sludge with different structures and colors, dark digested sludge and light lime sludge make the highest demands on the precise measurement of solids. Only SOLITAX sc solids probes with the patented dual scattered light method can satisfy these demands.
The SOLITAX sc sensor's self-cleaning wiper prevents erroneous values and maintenance problems by preventing biological growth. Gas bubbles in the sample do not interfere with the measurement.
All SOLITAX sc probes are factory calibrated in conformity with DIN EN ISO 7027 for long-term calibration stability. Calibration is not necessary. If the probe is used to measure the solids content, a one-point calibration suffices. Multipoint calibration is available for nonlinear samples.
Connects directly to all Hach sc controllers for immediate use. Helps improve process control and reduce treatment costs associated with polymer use, digester volume, and sludge handling. Any two SOLITAX sc sensors can be installed using one HACH sc200 Controller.
Suspended Solids probe, user manual
Accuracy:Turbidity up to 1000 NTU:
without calibration < 5% of the measured value ±0.01 NTU
with calibration < 1% of the measured value ±0.01 NTU
Cable Length:10 m (optional extension cables available)
Calibration Method:Turbidity: Formazin or STABLCAL Standard (at 800 NTU).
Suspended Solids: Sample specific, based on gravimetric TSS analysis with a correction factor procedure.
Flow:Max. 3 m/s (the presence of air bubbles affects the measurement)
Length:7.9 in ( 200 mm )
Maintenance Interval:1 h/month
Material:Optics Carrier and Sleeve: stainless steel 1.4571
Max. cable length Sensor-to-Controller:100
Measuring Principle:Infrared Duo scattered light technique for color-independent turbidity measurement. Turbidity in accordance with DIN EN 27027 / TS equivalent DIN 38414
Measuring Range TSS-Content:0.001 mg/L to 50 g/L
Measuring Range Turbidity:0.001 to 4000 NTU
Operating Temperature Range:0 to 40 °C
Pressure Limit:5 bar
Repeatability:ts content: < 3 %
Repeatability Turbidity:turbidity: < 1 %
Response Time T90:1 to 300 s adjustable
Weight Sensor:1.8 kg
- See more at: http://www.hach.com/solitax-turbidity-and-suspended-solids-ts-line-sc-stainless-steel-without-wiper-0-001-to-4000-ntu-0-001-mg-l-to-50-g-l-sensor-only-/product-details?id=7640286623#sthash.8Ku3fruQ.dpuf